Objective
To determine the diagnostic accuracy of different methods of blood pressure (BP) measurement compared with reference standards for the diagnosis of hypertension in patients with obesity with a large arm circumference.

Data sources
MEDLINE, EMBASE, Cochrane, DARE, Medion and Trip databases were searched.

Eligibility criteria
Cross-sectional, randomised and cohort studies of diagnostic test accuracy that compared any non-invasive BP tests (upper arm, forearm, wrist, finger) with an appropriate reference standard (invasive BP, correctly fitting upper arm cuff, ambulatory BP monitoring) in primary care were included.

Results
4037 potentially relevant papers were identified. 20 studies involving 26 different comparisons met the inclusion criteria. Individual patient data were available from 4 studies. No studies satisfied all QUADAS2 criteria. Compared with the reference test of invasive BP, a correctly fitting upper arm BP cuff had a sensitivity of 0.87 (0.79 to 0.93) and a specificity of 0.85 (0.64 to 0.95); insufficient evidence was available for other comparisons to invasive BP. Compared with the reference test of a correctly fitting upper arm cuff, BP measurement at the wrist had a sensitivity of 0.92 (0.64 to 0.99) and a specificity of 0.92 (0.85 to 0.87). Measurement with an incorrectly fitting standard cuff had a sensitivity of 0.73 (0.67 to 0.78) and a specificity of 0.76 (0.69 to 0.82). Measurement at the forearm had a sensitivity of 0.84 (0.71 to 0.92) and a specificity 0.75 of (0.66 to 0.83). Bland-Altman analysis of individual patient data from 3 studies comparing wrist and upper arm BP showed a mean difference of 0.46 mm Hg for systolic BP measurement and 2.2 mm Hg for diastolic BP measurement.

Conclusions
BP measurement with a correctly fitting upper arm cuff is sufficiently sensitive and specific to diagnose hypertension in patients with obesity with a large upper arm circumference. If a correctly fitting upper arm cuff cannot be applied, an incorrectly fitting standard size cuff should not be used and BP measurement at the wrist should be considered.

 

Irving, G; Holden, J; Stevens, R; McManus, RJ. (2016). Which cuff should I use? Indirect blood pressure measurement for the diagnosis of hypertension in patients with obesity: a diagnostic accuracy review. BMJ Open. 6, e012429

 